Station Setup
2300-RTD6

Station Summary

Station Programming dialog

After changes are made, click "Program" to write changes to station.

Settings Table

RTD Input Range 1 - PT100 (-200.0 to 850.0 C)
2 - Ni120 (-80.0 to 230.0 C)
3 - PT1000 (-200.0 to 850.0 C)
4 - Ni1000-DIN (-200.0 to 850.0 C)
5 - Ni1000-Landys&Gyr (-200 to 850.0 C)
6 - Ohms 10-400 (10.00-400.00) *
7 - Ohms 100-4000 (100.0-4000.0) *
RTD Display 1 - Deg C
2 - Deg F
COMS Reply Delay 0 to 990 mSec (20 mSec recommended)
Line Frequency 50 Hz
60 Hz

* Must be read into tags and scaled by a tag.

DIP Switch Settings

Switch 1-7 Address (See Address Table)
Switch 8 No Action, Leave OFF
Switch 9 Sensor Break -
OFF = RTD break = low value
ON = RTD break = high value
Switch 10 No Action, Leave OFF

LED Indications

1,2,3,4,5,6 INPUT LED OFF Input Normal
1,2,3,4,5,6 INPUT LED ON Input Faulty
P Flashing - Station is Powered up and working
R Flashes to indicate good modbus message arrived for this station
T Flashes to indicate reply message has been sent from this station

Station Info (Read from station)

30100 Dip Switch settings
30001 Station Type and Software Version (Type = 109)

Station Test Dialog:

Station Map Display

NOTE: For using Psuedo Digital inputs into a RTD6, Select the input range 1 - PT100 (-200.0 to 850.0 C), and the Sensor break must be ON, so that when the input contact is open, the RTD will read a high value.

Alternatively, the RTD Input Status can be used, for a good RTD reading "Healthy" will be returned and for a bad RTD reading "Faulty" will be returned.

Error checking can be provided by reading the RTD Input X Status value from the station as a digital, this is 1 (Healthy) when the RTD is reading OK, and 0 (faulty) when the probe is in error.