FAQ |
Readings from a 2100-M connected to a 2100-A16 are not correct |
If readings from a 2100-M connected to a 2100-A16 are not correct, check the
clock and reset lines from the A16 to 2100-M are wired correctly.
The clock LED will flash at a much faster rate than the reset LED.
If this fault is present, all probes on the 2100-M will read the same value.